Product
Features
Product Information
Size / Weight
Operating Temperature
Display
Printer
Interface
Power
Built-in battery
Cartridge
i-Smart 300 Cartridge
Measurement items
Calculation items
Sample type and volume
Testing time
Calibration
Cartridge specifications
100 tests for 2 to 3 weeks
200 tests for 2 to 3 weeks
300 tests for 2 to 3 weeksExpiration date
Quality Control
i-Smart QC
Features
Use with
Level
Volume
i-Smart 300 Blood Gas Analyzer
257 X 394 X 252 mm / 7.1 kg
15-35 °C
8 inch color LCD (touchscreen)
Two-inch (built-in) thermal printer
Serial (RS-232), LAN (RJ45 Ethernet)
100–240 VAC, 50/60 Hz
Up to 2 hours of use (lithium-ion cell)
pH, pCO2, pO2, Na+, K+, Cl–, Ca2+, Hct
pH(T), pCO2(T), pO2(T), ctCO2, cHCO3-, cHCO3-(std), BE(ecf), BE(B), ctHb, Ca2+(7.4), pO2(A-a), sO2
Heparinized whole blood/100 μL
50 seconds
Automatic or manual
50 tests for 2 to 3 weeks
6 months from the manufacturing date
Blood gas Quality control material
i-SmartCare 10, i-Smart 300, and i-Smart 300 VET Analyzers
1, 2, 3
2.5 mL * 10 vials per level